Biosystems Bts 350 User Manual //free\\ Jun 2026
BTS-350 Manual de Usuario: Guía Completa para ... - Studocu
Used for tests like Glucose, Cholesterol, and Total Protein.
Hold the tube steady until the peristaltic pump stops drawing liquid, then remove the vessel immediately to prevent air bubbles from entering the flow cell. biosystems bts 350 user manual
Standard configuration includes 8 filters: 340, 405, 505, 535, 560, 600, 635, and 670 nm. Bandwidth: ≤is less than or equal to Photometric Range: 0.0 to 3.5 Absorbance (A). Fluidics and Temperature Control
Rerun the water blank, recalibrate the assay, and test the control sample a second time. BTS-350 Manual de Usuario: Guía Completa para
: Select "Concentration," choose your programmed test (e.g., Glucose, Albumin), and follow the on-screen prompts to insert a baseline (blank) followed by your sample. Key Technical Features Optical System
Clean the exterior surface using an alcohol wipe to remove biological residue. Peristaltic Pump Standard configuration includes 8 filters: 340, 405, 505,
Place the instrument on a flat, stable, and vibration-free surface.
: Utilizes 8 high-intensity LEDs (340, 405, 505, 535, 560, 600, 635, and 670 nm) plus one free position for optional filters. These LEDs provide a practically unlimited lifetime and low energy consumption. Flexible Measurement Modes : Supports multiple analysis methods, including: End-point (monochromatic/bichromatic) Kinetics and Fixed Time Absorbance and Differential Mode Precise Fluidic & Temperature Control Sipping Volume