Diagnostica Stago Start 4 Coagulation Analyzer Manual Jun 2026
Reference: User Manual, Section 5.3
: A built-in 40-column thermal printer for immediate result recording. Core Functionality and Test Menu diagnostica stago start 4 coagulation analyzer manual
Open the printer door, clear any torn paper fragments, reload the thermal paper roll straight, and feed it through the slot cleanly. Poor manual pipetting technique or contaminated cuvettes. Reference: User Manual, Section 5
Unlike optical analyzers that measure changes in light transmission, the Start 4 utilizes an electromagnetic mechanical detection system. This eliminates optical interference caused by hemolyzed, icteric, or lipemic plasma. How the Mechanical System Works Unlike optical analyzers that measure changes in light
: A steel ball is dispensed into each cuvette. The analyzer measures the change in sample viscosity by tracking this ball's oscillation between magnetic transducers.
: Use a 1:10 dilution of 5% sodium hypochlorite to clean, then rinse and dry with a slightly moistened paper tissue.
: Place cuvettes with samples into the incubation area and start the corresponding timer (A–D). Measurement : Transfer the cuvette to a measurement channel.