Simco Fmx002 Manual Top File
Before taking a measurement, point the sensor away from any charged surfaces and press the button. This ensures your baseline is neutral. Step 2: The Distance Rule (25mm) The is calibrated for a specific distance: exactly ( ) from the target surface.
Turn off immediately. Clean with compressed air or send for service. Battery voltage is low Replace the 9V battery immediately to maintain accuracy. Inconsistent distance readings LED range guides are misaligned
The oscillating chopper sensor at the top of the meter is highly sensitive.
Do not use the meter in areas with high humidity (>60% RH) or extreme temperatures.
Mount the FMX-002 securely on a non-conductive fixture exactly 1 inch away from the plate. Ground the FMX-002 housing. The Verification Process simco fmx002 manual top
: Do not zero the meter in free air unless the air is known to be ion-free. Free air can have a residual field of 1-2 kV due to ambient ions.
: Ensure the instrument is connected to the earth using the supplied grounding lead for accurate readings. www.joongang21.com Maintenance Tips
: Freezes the current reading on the display, which is useful when measuring in hard-to-see areas. MODE / Ion Balance
Note that the zero button will not function if the device is in HOLD mode or if the current reading is already too high, indicating the sensor is already near a strong charge. 3. Understanding the Display Before taking a measurement, point the sensor away
The sensor is delicate; avoid dropping the unit.
+---------------------------------+ | [ LCD DISPLAY ] | | (Bar Graph & Digital Volts) | +---------------------------------+ | [ZERO] [HOLD] [IB] [POWER]| | (Gray) (Green) (Blue) (Red) | +---------------------------------+ | [ SENSOR HEAD ] | | (Ranging Focus LEDs Inside) | +---------------------------------+
To execute standard field mapping on surfaces, materials, or assembly lines, follow these sequential steps: Simco FMX-002 Electrostatic Fieldmeter Calibration
Conductive resin body ensuring proper path-to-ground. 🧭 Device Interface and Control Layout Turn off immediately
This overlapping point indicates exactly 1 inch of separation. Step 4: Reading and Holding Data
This comprehensive technical guide details the setup, key features, operation, and maintenance procedures outlined in the official user manual. 🛠️ Key Technical Specifications
Static electricity can be a silent disruptor in industrial environments, causing everything from minor shocks to catastrophic equipment failure. To manage it, you need precision tools. Enter the Simco FMX-002 Electrostatic Fieldmeter Go to product viewer dialog for this item.
You can find detailed calibration services and support via Micro Precision or consult the official Simco-Ion documentation. Simco FMX-002 Electrostatic Fieldmeter Calibration
Freezes the current measurement on the screen. Ideal for logging data in tight spaces. The Display Panel